KARL STORZ SE & CO. KG ASM,HEAD,IMAGE 1 S, 4U Back to Search Results
Model Number TH120
Medical Device Problem Codes Display or Visual Feedback Problem (1184); No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the camera presented a shaking image and sometime a black screen.The shaking is not permanent.It occured few times the sound of the units are very acute and is suggesting a fan problem.This situation occured during use and during a test in our technical department.The procedure was completed successfully with no surgical delay or prolongation of surgery.No negative impact in state of health reported.Cross reference: (b)(4).
 
Additional Manufacturer Narrative
The device in question was returned to the manufacturer.The evaluation is anticipated, but not yet begun.The investigation will be performed by a designated karl storz employee.The event is filed under internal karl storz complaint id: (b)(4).
 
Additional Manufacturer Narrative
Device evaluation: during the manufacturer's technical inspection of the returned devices, the error description provided by the customer, "shaking image /black screen sometimes" could not be confirmed.Even though the 3 associated units, tc201, tc304 and th120 were tested in the system in continuous operation over several days, the described failure of the customer could not be detected/ determined.The most probable root cause therefore is the cabling on site which may have a loose or not correct connected connection.This event is filed under internal complaint id (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASM,HEAD,IMAGE 1 S, 4U
Common Device Name
ASM,HEAD,IMAGE 1 S, 4U
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
anja fair
2151 e grand ave
el segundo, CA 90245
MDR Report Key22589313
Report Number9610617-2025-01289
Device Sequence Number18393569
Product Code FET
UDI-Device Identifier04048551370506
UDI-Public04048551370506
Combination Product (Y/N)N
Initial Reporter CountryTS
PMA/510(K) Number
K131953
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTH120
Device Catalogue NumberTH120
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/04/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2025
Supplement Date Received by Manufacturer12/02/2025
Initial Report FDA Received Date07/24/2025
Supplement Report FDA Received Date12/04/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-