| Model Number |
TH120 |
| Medical Device Problem Codes |
Display or Visual Feedback Problem (1184); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the camera presented a shaking image and sometime a black screen.The shaking is not permanent.It occured few times the sound of the units are very acute and is suggesting a fan problem.This situation occured during use and during a test in our technical department.The procedure was completed successfully with no surgical delay or prolongation of surgery.No negative impact in state of health reported.Cross reference: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The device in question was returned to the manufacturer.The evaluation is anticipated, but not yet begun.The investigation will be performed by a designated karl storz employee.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: during the manufacturer's technical inspection of the returned devices, the error description provided by the customer, "shaking image /black screen sometimes" could not be confirmed.Even though the 3 associated units, tc201, tc304 and th120 were tested in the system in continuous operation over several days, the described failure of the customer could not be detected/ determined.The most probable root cause therefore is the cabling on site which may have a loose or not correct connected connection.This event is filed under internal complaint id (b)(4).
|
| |
|