| Model Number |
359067 |
| Medical Device Problem Codes |
High impedance (1291); Ambient Noise Problem (2877)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Event or Problem Description
|
|
Out of range shock impedance (greater than 150 ohms) and noise on the rv channel were seen in recordings transmitted to home monitoring.Lead to be evaluated further and remains implanted.Should additional information be received this file will be updated.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.The device is currently not available for analysis.No conclusion can be drawn based on available information at the moment.The file is closed.The investigation will be re-opened should additional data become available.
|
| |
|
Search Alerts/Recalls
|