| Catalog Number |
0233050087 |
| Medical Device Problem Codes |
Optical Problem (3001); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/26/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
The adapter was not recognized and the light source stopped working.The permanent magnet had lost its magnetic force.Therefore, there was loss of light.
|
| |
|
Event or Problem Description
|
|
The adapter was not recognized and the light source stopped working.The permanent magnet had lost its magnetic force.Therefore, there was loss of light.
|
| |
|
Additional Manufacturer Narrative
|
|
Multiple attempts have been made to retrieve the device for evaluation, and it is yet to be received in-house.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: no light output probable root cause: broken optical fibers, poor light cable alignment in jaw, residue on fiber tip, nonuniform light output intermittent electrical component contact in safelight circuit, reed switch assembly malfunction magnet missing from safelight adapter use error the reported failure mode will be monitored for future reoccurrence.D4 lot # was updated.
|
| |
|