• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL ENVOY; PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL ENVOY; PERCUTANEOUS CATHETER Back to Search Results
Catalog Number 67125895DB
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/11/2025
Type of Reportable Event Malfunction
Event or Problem Description
The healthcare professional reported that during a procedure on (b)(6) 2025, the hub of the envoy da xb, 6f, 95cm, mpd (67125895db / 31502173) was broken when it was first opened.The physician replaced it with another envoy da xb, 6f, 95cm, mpd (67125895db) and the procedure was successfully completed.No negative impact to the patient.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.(b)(4).Information regarding patient identifier, date of birth, age, sex, gender, weight, race, and ethnicity were not provided.Section e.1: the initial reporter phone is not available / reported.Section h.3: the device is available to be returned for evaluation and testing.However, it has not been received to date.If the device returns, a device investigation will be performed.A review of manufacturing documentation associated with this lot (31502173) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no internal actions related to device manufacture or inspection.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by johnson & johnson medtech, or its employees that the report constitutes an admission that the product, johnson & johnson medtech, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report that the product analysis lab received the complaint device on 26-aug-2025.A supplemental 3500a report will be submitted once the product investigation has been completed.Updated sections: b.4, d.9, g.3, g.6.H.2, h.3, h.6, and h.11.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report the investigational finding of the returned device.The complaint product was returned and received for evaluation and analysis.The investigation is documented below.Investigation summary: a non-sterile envoy da xb, 6f, 95cm, mpd was received contained in the decontamination pouch.Visual inspection was performed.The luer hub was observed to be cracked at the fusion area.The issue reported in the complaint related to the hub of the envoy catheter being broken was confirmed during the inspection of the returned device.Hub damage during attachment/removal of hemostasis valve is a potential failure mode that can result from the handling of hemostasis valve and hub where excessive force may have been applied, causing device damage.Devices undergo 100% inspection for hub damage during hub injection molding process.Thus, it is not likely that the device left the facility in a damaged condition; intra-operative factors may have contributed to the issue reported, however, with the amount of information available, a root cause cannot be established.A review of manufacturing documentation associated with this lot (31502173) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no internal actions related to device manufacture or inspection.As part of johnson & johnson medtech quality process, all devices are manufactured, inspected, and released to approved specifications.It should be noted that product failure could be caused by multiple factors.The instruction for use (ifu) contains the following caution: ¿ do not use a catheter that has been damaged in any way.If damage is detected, replace with another envoy guiding catheter that is not damaged.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.Updated sections: b.4, g.3, g.6.H.2, h.3, h.6, and h.11.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENVOY
Common Device Name
PERCUTANEOUS CATHETER
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CERENOVUS INC. (JUAREZ)
circuito interior norte #1820
parque industrial salvarcar
juarez chihuahua 32574
MX   32574
Manufacturer Contact
henry boden
31 technology dr
irvine, CA 92618
949788-2078
MDR Report Key22599541
Report Number3008114965-2025-00720
Device Sequence Number5016754
Product Code DQY
UDI-Device Identifier10886704075110
UDI-Public(01)10886704075110(17)270930(10)31502173
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K000715
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number67125895DB
Device Lot Number31502173
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/09/2025
Supplement Date Received by Manufacturer08/26/2025
11/07/2025
Initial Report FDA Received Date07/24/2025
Supplement Report FDA Received Date08/26/2025
11/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/30/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-