| Model Number |
BF-P180 |
| Medical Device Problem Codes |
Material Discolored (1170); Material Deformation (2976); Sharp Edges (4013)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/26/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: probable causes due to some kind of stress such as damage or deterioration of parts, electrical problem, environmental temperature problem, and maintenance problem.The most probable cause of this complaint is expected or random component failure without any design or manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the bronchovideoscope exhibited damage on bending tube, metal part is sticking out from breakage on a-rubber) and corrosion on the connecting tube.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|