• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SENSIA IPG; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SENSIA IPG; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number SEDR01
Medical Device Problem Codes Premature Discharge of Battery (1057); Communication or Transmission Problem (2896); Reset Problem (3019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Continuation of d10: 4068-45 lead implanted on (b)(6) 1997.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that one of the pacing leads exhibited out of range (oor) impedances and a polarity switch.It was also reported that an electrical reset occurred and there was a possible telemetry issue on the implantable pulse generator (ipg).The ipg was removed and replaced.The right ventricular (rv) lead remains in use.No patient complications have been reported as a result of this event.
 
Event or Problem Description
It was further reported that the ipg was explanted due to battery depletion.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Product event summary: the device was returned and analyzed.Analysis of the device revealed normal battery depletion.The device was returned after an 108 months for premature depletion and electrical reset.The device was received in a no telemetry no output condition.The device loss of telemetry and all functions were due to a severely depleted battery.The device was fully functional and operated under normal current drain when powered with an external supply and no hybrid anomalies were observed.The device passed longevity.It was determined that the device met specifications for normal operation.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSIA IPG
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key22606678
Report Number3004209178-2025-12683
Device Sequence Number5494971
Product Code NVZ
UDI-Device Identifier00613994222138
UDI-Public(01)00613994222138(17)170728
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P980035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/28/2017
Device Model NumberSEDR01
Device Catalogue NumberSEDR01
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2025
Supplement Date Received by Manufacturer09/12/2025
11/10/2025
Initial Report FDA Received Date07/25/2025
Supplement Report FDA Received Date09/24/2025
11/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/02/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5076-52 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-