Additional information provided in h.6.And h.11.A sample was not received at the manufacturing site; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The photo attached to the complaint was reviewed by the photo 2 shows pack lot number.Reported issue confirmed from photo 1.The investigation manufacturing site.The photo1 shows bent and deformed phacoemulsification tip and conducted a non-conformance review of the reported lot number.No deviations were identified and all corresponding production release specifications defined in the device master record were met.The evaluation of the attached photo confirms the balance tip bent and deformed as noted by the customer.Based on the evaluation of the information, since the sample was not received at the manufacturing site, how and when the balance tip was bent and deformed was unable to identify the root cause or origin of the reported event.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.All phacoemulsification tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|