The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
The reported event is confirmed manufacturing related.This is addressed by capa 12182448.Visual evaluation noted received one (1) used all-silicone foley catheter with sample port connector, syringe and packaging label.Visual inspection noted that there was a hole in the drainage funnel measuring (0.050").Therefore, the reported event is confirmed and does not meet specifications per ip7603444 rev 0 which states catheter leaks, valve leaks, balloon perforation, lumen to lumen, prematurely deflated and dislodged balloon are not allowed, and must inflate and deflate with the use of a syringe.The root cause for this failure, per capa 12182448, is the silicone injected into the forming cavity is entering pre-cured, which causes overheating in the gate area or injection point, as a result, the breaking point between the gate and the molded part is not properly generated, which prevents automatic separation.This requires manual detachment, increasing the risk of perforation at the adapter injection point due to the high temperature in the gate caused by the lack of flow in the jacket.Risk review, labeling review, and dhr review are not required as event has already being investigated per capa 12182448.Corrections: d, e, f, h.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|