| Model Number |
P20263-1020G |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Fluid Discharge (2686)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient experienced fluid collection and wound drainage approximately one month after a cosmetic revision procedure performed with breast implants and ovitex prs.The surgeon performed a washout to address the drainage issue.
|
| |
|
Additional Manufacturer Narrative
|
|
A tela bio medical affairs spoke with the reporting surgeon during the week of july 7th, 2025 and all information reasonably known to tela bio from this communication and all other investigation is included in this report.The surgeon reported four similar cases simultaneously that occurred intermittently over the course of approximately six months and did not offer full case details for each event.A batch record review was conducted which showed no non-conformances or anomalies.The lot met all release specifications and was sterilized utilizing validated sterilization parameters.Wound complications and fluid collection are known risks associated with soft tissue reinforcement procedures.
|
| |
|
Search Alerts/Recalls
|