Catalog Number 110031426
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Implant Pain (4561)
Date of Event 01/18/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: - concomitant medical products, part (lot): associated product information: -110031400 (64263046).Associated product information, part (lot).-211218 (125640).-115313 (624810).-211244 (739250).-110027734 (136100).-211228 (480440).-211228 (480440).Multiple mdr reports were filed for this event.Associated reports are available for the applicable concomitant products below in the d10 narrative.The customer has indicated that the product will not be returned to zimmer biomet for investigation as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that a patient entered clinical study upon revision of right rtsa with repair of commuted humeral fracture.Subsequently, the patient experienced ongoing pain with no identified implant abnormalities and received injection approximately 4 years post-implantation with resolution.Remains implanted at study completion.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: b4; b5; d2; d9; g1; g3; g6; h1; h2; h3; h4; h6; h10.The reported event is not confirmed.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Crfs were provided and reviewed by a health care professional.Review of the available records identified the following: patient reported persistent pain and muscle tenderness near cable fasteners.Xray/ct showed no hardware issues; ot prescribed.Patient again reported ongoing pain, xray identified no hardware issues, trigger point injection given.Mri ¿ negative, symptoms gradually resolved without intervention.Study completion ¿ device remains implanted.A definitive root cause cannot be determined.If any further information is found that would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPREHENSIVE REVERSE SHOULDER SYSTEM VIVACIT-E HIGHLY CROSSLINKED POLYETHYLENE
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key22610508
Report Number0001822565-2025-02664
Device Sequence Number5384866
Product Code PHX
UDI-Device Identifier00887868231575
UDI-Public(01)00887868231575(17)240229(10)64291230
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K181611
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date02/29/2024
Device Catalogue Number110031426
Device Lot Number64291230
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/27/2025
Supplement Date Received by Manufacturer12/05/2025
Initial Report FDA Received Date07/25/2025
Supplement Report FDA Received Date12/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/19/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age83 YR
Patient SexFemale
Patient Weight68 KG
Patient RaceWhite
-
-