• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. COMPREHENSIVE VRS GLENOID; SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. COMPREHENSIVE VRS GLENOID; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Catalog Number 110027734
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Implant Pain (4561)
Date of Event 01/18/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: concomitant medical products, part (lot): associated product information: -211218 (125640).-115313 (624810).-211244 (739250).-211228 (480440).-211228 (480440).Associated product information, part (lot) -110031400 (64263046).-110031426 (64291230).Multiple mdr reports were filed for this event.Associated reports are available for the applicable concomitant products below in the d10 narrative.The customer has indicated that the product will not be returned to zimmer biomet for investigation as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that a patient entered clinical study upon revision of right rtsa with repair of commuted humeral fracture.Subsequently, the patient experienced ongoing pain with no identified implant abnormalities and received injection approximately 4 years post-implantation with resolution.Remains implanted at study completion.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: b4; b5; d2; e1; e2; e3; g1; g3; g6; h1; h2; h3; h4; h6; h10.The reported event is not confirmed.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing.Crfs were provided and reviewed by a health care professional.Review of the available records identified the following: persistent pain and muscle tenderness near cable fasteners.Xray/ct showed no hardware issues ; ot prescribed.Again ongoing pain, xray ¿ no hardware issues.Trigger point injection given.Mri ¿ negative.Symptoms gradually resolved without intervention.Study completion; remains implanted.A definitive root cause cannot be determined.If any further information is found that would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPREHENSIVE VRS GLENOID
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key22610545
Report Number0001825034-2025-02246
Device Sequence Number18542706
Product Code PHX
UDI-Device Identifier00880304818934
UDI-Public(01)00880304818934(17)191213(10)136100
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K152754
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date12/13/2019
Device Catalogue Number110027734
Device Lot Number136100
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/27/2025
Supplement Date Received by Manufacturer11/03/2025
Initial Report FDA Received Date07/25/2025
Supplement Report FDA Received Date12/02/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/13/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age83 YR
Patient SexFemale
Patient Weight68 KG
Patient RaceWhite
-
-