• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH TELESCOPE "TRUEVIEW II", 2.7MM, 30°, SHORT, AUTOCLAVABLE; HW70963

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH TELESCOPE "TRUEVIEW II", 2.7MM, 30°, SHORT, AUTOCLAVABLE; HW70963 Back to Search Results
Model Number A70963A
Medical Device Problem Codes Fracture (1260); Optical Problem (3001)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/29/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was observed that during the device evaluation, the rigid scope exhibited a breakage on objective cover glass, insertion tube broken and light guide fiber broken.There was no patient involvement.
 
Additional Manufacturer Narrative
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: cause traced to component failure of the objective cover glass and light guide fiber.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE "TRUEVIEW II", 2.7MM, 30°, SHORT, AUTOCLAVABLE
Common Device Name
HW70963
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key22610569
Report Number9610773-2025-04725
Device Sequence Number14798755
Product Code HRX
UDI-Device Identifier04042761010658
UDI-Public04042761010658
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K951354
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberA70963A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/29/2025
Initial Report FDA Received Date07/25/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/19/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-