| Model Number |
VTICMO 12.6 (-16.50/1.5) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/27/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced lens rotation not associated to a low vault.Lens was repositioned but that did not resolve the problem.The lens exchanged for a longer length lens and the problem was resolved.Cause of the event was reported as unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Manufacture narrative: secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim #: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3 - device evaluation: the lens was returned in a microcentrifuge vial dry.Visual inspection found no visible damage to lens.Dimensional inspection found the lens to be within specifications.Claim #(b)(4).
|
| |
|