HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0; VENTRICULAR (ASSISST) BYPASS
|
Back to Search Results |
|
| Model Number |
1420-CONTROLLER |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation of this event is pending and a supplemental report will be sent upon its completion.Continuation of d10: dvpa2d4 icd implant date: on (b)(6) 2023, 7122q lead implant date: on (b)(6) 2014.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a pioneer controller exhibited a controller fault alarm.The alarm was caused by the internal battery reaching end of life.After the audio was permanently silenced, there was no single beep when a power source was added.The controller remains in use.No patient symptoms, complications, or injuries were reported as associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A supplemental report is being submitted for device evaluation and investigation completion.Product event summary: the controller was not returned for evaluation.A review of the device history record was not performed as the reported event and analysis are not related to a manufacturing or servicing issue.Review of the controller log files revealed two (2) controller fault alarms logged on (b)(6) 2025 at 16:15:39 and at 17:32:48, as well as multiple event exceptions logged on (b)(6) 2025, indicating an issue with the internal battery.In addition, log files revealed that the controller had been in use for more than two (2) years.As a result, the reported event was confirmed.Applicable risk documentation, experience with events of similar circumstances, and the available information were considered; possible root causes of the reported controller fault alarm event may be attributed, but not limited, to a reduced charge capacity of the internal battery and/or a faulty internal battery charger integrated circuit.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|