W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Catalog Number |
CXT231412 |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Code |
Obstruction/Occlusion (2422)
|
| Date of Event |
07/03/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
The following information was reported to gore: on (b)(6) 2025, the patient underwent endovascular treatment for the abdominal aortic aneurysm using the gore® excluder® conformable aaa endoprosthesis, the gore® excluder® aaa endoprosthesis, and the gore® excluder® iliac branch endoprosthesis.After contralateral cannulation, resistance was encountered when elevating the gore® dryseal flex introducer sheath.Following deployment of the contralateral limb, it was confirmed that the trunk-ipsilateral leg endoprosthesis had unintentionally covered approximately 80% of the right renal artery.A stent was placed in the right renal artery. resistance was encountered when elevating the sheath after contralateral cannulation.It is suspected that the main body may have been unintentionally lifted along with the sheath due to the application of excessive force.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Search Alerts/Recalls
|
|
|