| Catalog Number |
EIC8898-01 |
| Medical Device Problem Codes |
Degraded (1153); Melted (1385); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that during an ent surgery, the active electrode of two (2) procise max coblator wands was getting damage very quickly, the electrodes were melted.The surgery was completed with a smith and nephew back up device.There was a delay greater than 30 minutes.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).H11 h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation.However, an image evaluation was performed and found the wand tip with wear and debris.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause of the reported event could not be determined since findings cannot lead to a clear cause of the reported event.Factors that could have contributed to the reported event include 1) hitting the device tip against a hard surface.2) using the device as a lever to enlarge surgical site.3) mechanical displacement of tissue through applied force.4) activating the device above recommended settings for prolonged periods of time.Based on this investigation, the need for corrective action is not indicated.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: h3, h6: the reported device was received for evaluation.A visual inspection revealed the returned instrument shows no manufacturing abnormalities.Product was out of the original packaging.No packaging returned.Two wands were received.There is no way to determine one from another.Wand 1 the tip is worn from use.The electrode is thin from use.The electrode is missing a portion of its body and is bent from use.Wand 2 the tip is worn from use.The electrode is not damaged on this wand.A functional evaluation was performed, the devices were tested and plugged into the controller and registered settings (7,3).The wands can produce plasma with their remaining electrodes and have no issues activating coag.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A definitive root cause for the reported issue could not be determined.Factors that could have contributed to the failure include procedural variance, or deviations from established protocols, increasing risks and leading to unintended consequences (1) using the wand above default output settings, thus causing the electrodes to become eroded extensively.2) pressing the tip against hard or bony surfaces).Based on this investigation, the need for corrective action is not indicated.Based on this investigation, the need for corrective action is not indicated.
|
| |
|