Two samples of product number 11171447 were submitted for quality notification.The two samples were connected to bd products 1001224-0500 and 10014881.The samples were visually inspected for damages or abnormalities, and none were found.The samples were primed, and leakage was noted at the union between the texium connectors, product 1101224-0500, and a y-site smartsite on product 10014881, but no leakage was identified from product 11171447.The leakage of the texium product will be addressed in bd complaint (b)(4).The customer complaint of leakage for product 11171447 was not confirmed.A root cause was not investigated because the reported product, 11171447, did not leak.A device history record review for model 11171447, lot number 24115166 was performed.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.
|