MAUDE Adverse Event Report: SOL-MILLENNIUM MEDICAL INC MONOJECT; MONOJECT HYPODERMIC SAFETY NEEDLE 25G*1 1/2"

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOL-MILLENNIUM MEDICAL INC MONOJECT; MONOJECT HYPODERMIC SAFETY NEEDLE 25G*1 1/2" Back to Search Results
Model Number 1182515
Medical Device Problem Code Partial Blockage (1065)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/27/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There is no evidence of quality issues associated with this complaint based on the archive sample analysis.Based on the determined risk priority number, the residual risk related to the complaint is considered acceptable.Additionally, no trends related to this issue have been identified during our track-and-trend analysis.Sol millennium continues to monitor this kind of issue.Additional assessment will be taken if a significant pattern emerges.Unconfirmed.
 
Event or Problem Description
Customer reported that when being utilized it is difficult to push medications through, at times unable to push medications or feels as if getting push back.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONOJECT
Common Device Name
MONOJECT HYPODERMIC SAFETY NEEDLE 25G*1 1/2"
Manufacturer (Section D)
SOL-MILLENNIUM MEDICAL INC
311 s. wacker dr
suite 4100
chicago IL 60606
Manufacturer (Section G)
ZHEJIANG KINDLY MEDICAL DEVICES CO., LTD.
no. 758, 5th binhai road
binhai industrial park
longwan ,wenzhou,, zhejiang province
CH  
Manufacturer Contact
mohanapriya kannupaiyan
311 s. wacker dr
suite 4100
chicago, IL 
MDR Report Key22628551
Report Number3014312726-2025-00173
Device Sequence Number8155335
Product Code FMI
UDI-Device Identifier10884521544697
UDI-Public(01)10884521544697(10)02309056
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K123684
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1182515
Device Lot Number02309056
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/03/2025
Initial Report FDA Received Date07/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-