| Catalog Number |
8065751114 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Firing Problem (4011)
|
| Health Effect - Clinical Code |
Retinal Tear (2050)
|
| Date of Event |
07/17/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A healthcare professional reported that an ophthalmic laser probe pierced the retina while the power was being reduced during a vitrectomy procedure.The surgery was completed on the same day, and the patient's current condition was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information is provided in sections b.5 and h.6.The manufacturer internal reference number is:(b)(4).
|
| |
|
Event or Problem Description
|
|
Additional information received indicated that the ophthalmic laser probe had been too powerful.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information is provided in sections h.6 and h.11.The product under investigation is not a serviceable device.Therefore, a service record review was not performed.There was no product returned on this investigation.Based on the information obtained, the root cause of the reported event is inconclusive.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.Based on the information obtained, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
| |
|