|
SYNTHES GMBH MINI QUICK COUPLING; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
532.011 |
| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H11 additional narrative: device evaluation: this device was returned for evaluation.A visual and functional assessment was performed which determined that the device passed all pre-repair diagnostic assessments and no failures were identified.Therefore, the reported condition was not confirmed, and an assignable root cause was not determined.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H11 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
|
| |
|
Event or Problem Description
|
|
Report 3 of 3 for (b)(4).It was reported that the quick coupling device could not hold the sagittal saw attachment device and the reaming attachment device and detached during use.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2025.All available information has been disclosed.
|
| |
|
Search Alerts/Recalls
|
|
|