MEDTRONIC SOFAMOR DANEK USA, INC LONGITUDE II; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 5584099
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/08/2025
Type of Reportable Event Malfunction
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having revision surgery for the the removal of the existing l5 tsrh rp screw due to l5/s adjacent intervertebral disorder, l3/s fixation was performed in connection with l3-5.Followed by the insertion of a new solera screw on l5/s transforaminal lumbar interbody fusion (tlif) was performed.It was reported that the procedure encountered challenges due to pedicle hardening, causing the tip of the driver to break during screw insertion.Replacement driver were used, but that replacement driver also broke during screw removal.The screw size was adjusted from 7.5mm to 6.5mm, and the screw was eventually extracted using a short rod and reverse rotation with a power grip.These issues caused a delay of less than 60 minutes in the overall procedure time.There was no patient symptom reported.There were no further complications reported regarding the event.The allegation on screw was the rod piece and the screw were stuck on the distal side of the product.The damaged screwdriver tip remains in the screw crown.Since it was not possible to remove or insert with a normal screwdriver, it was fastened once with a short rod and grasped the rod with a rod gripper and removed.There was no malfunction against tsrh rp1 (unknown tsrh rp).Due to adjacent intervertebral disorder, the nail was pulled out to extend the fixation.No patient symptoms.The product was discarded in the hospital.
 
Additional Manufacturer Narrative
E1: first name and last name of initial reporter is unknown.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3: product analysis of product:(b)(4), lot:sw19h018 visual and optical examination identified that the part of the tip of the instrument has been broken off.The metal deformation gives indication that the failure was the result of overload.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LONGITUDE II
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key22629162
Report Number1030489-2025-02514
Device Sequence Number7994329
Product Code LXH
UDI-Device Identifier00643169163065
UDI-Public(01)00643169163065(11)191022(10)SW19H018
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5584099
Device Catalogue Number5584099
Device Lot NumberSW19H018
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2025
Supplement Date Received by Manufacturer09/16/2025
Initial Report FDA Received Date07/28/2025
Supplement Report FDA Received Date10/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/22/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-