| Catalog Number |
515086 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the needle popped out of injector.Event description: when i tried to remove the injector after dosing, the needle popped out.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow mdr for correction, see d tab.Medical device catalog 515086.Medical device lot 2407013.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) follow up mdr for device evaluation: one sample was provided to our quality team for investigation.Through visual inspection, the hook of the safety sleeve is observed to be broken the injector needle was exposed, verifying the reported incident.A device history review was performed for reported lot 2407013, no deviations or non-conformances were identified during the manufacturing process that could have contributed to this observation.Product undergoes inspections throughout the manufacturing process to ensure the quality and functionality of the device.Testing results were reviewed for the reported lot and found all product met required specifications, no issues related to this incident were found.It is important to grip the white injector components when engaging/ disengaging; do not grip the blue safety sleeve.If the safety sleeve grips become dislocated, the injector is activated causing needle exposure.To prevent damage to the handles of the safety sleeve, the injector must be removed by pulling it backwards and must not be forcefully engaged.The instructions for use must be carefully followed when using phaseal devices in order to avoid any damage to the product that may result in the device not functioning as intended.Based on our sample evaluation, it was determined that needle became exposed due to the connection/disconnection of the device.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|