• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number 9-ASD-017
Medical Device Problem Code Difficult to Fold, Unfold or Collapse (1254)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/03/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2025, a 17mm amplatzer septal occluder was chosen for implant using a 10f amplatzer trevisio intravascular delivery system.During procedure, a cobra bulging on the left-atrial disc was noted.The deformed device was never released from the delivery cable while inside the patient.A new 18mm amplatzer septal occluder was chosen and completed the procedure.There was no report of any interaction with cardiac structures during deployment and no report of any angulation/kink noticed with the delivery system.There was no reported adverse consequences to the patient.
 
Additional Manufacturer Narrative
An event of device deformity could not be confirmed.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all defined manufacturing specifications.Additionally, the lot was reviewed for similar complaints, and there is no indication of a lot-specific product issue.The cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.Please note that per the instructions for use, the recommended size delivery system for use with a 17 mm amplatzer septal occluder is an 7f.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key22631731
Report Number2135147-2025-04145
Device Sequence Number9002581
Product Code MLV
UDI-Device Identifier00811806010144
UDI-Public(01)00811806010144(17)261031(10)8219791
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
P000039
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9-ASD-017
Device Catalogue Number9-ASD-017
Device Lot Number8219791
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/03/2025
Supplement Date Received by Manufacturer09/01/2025
Initial Report FDA Received Date07/28/2025
Supplement Report FDA Received Date09/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
10F AMPLATZER TREVISIO INTRAVASCULAR DELIVERY SYST
Patient SexUnknown
-
-