| Catalog Number |
STP-PR-001 |
| Medical Device Problem Code |
Low Readings (2460)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/26/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that an inaccuracy between the continuous glucose monitor (cgm) and the blood glucose (bg) meter occurred.The sensor was inserted into the arm which is off-label usage of the device on (b)(6) 2025.Data was evaluated and the allegation was confirmed.The probable cause could not be determined via data.The reported glucose values fall within the d zone of the parkes error grid.The data investigation found a difference in values that falls within the b zone of the parkes error grid.No injury or medical intervention was reported.
|
| |
|