• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH GAMMA3 - LONG NAIL KIT R1.5, TI, LEFT D11X420MM X 125DEG; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH GAMMA3 - LONG NAIL KIT R1.5, TI, LEFT D11X420MM X 125DEG; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 35251420S
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/01/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Once the investigation is completed, any additional information will be reported in a supplemental report.
 
Event or Problem Description
It was reported that after 6 month, the nail was removed since it broke off at the cephalic screw part.
 
Event or Problem Description
It was reported that after 6 month, the nail was removed since it broke off at the cephalic screw part.
 
Additional Manufacturer Narrative
The reported event could be confirmed since evaluation matches the reported failure mode.In the case presented an unknown patient was initially treated on (b)(6) 2024 with above shown gamma3 - long nail, which was noticed to be broken as per event details given ¿6 months after the surgery in which the nail was inserted¿ the received nail is completely broken in the webs of the proximal lag screw thru hole.According to the damage and the evidence of the breakage surfaces, the nail broke in a fatigue fracture.Appearance identified a high axial load application to the implants which is also supported by signs of damage verified at one locking screw.Areas with signs of friction / seizing (bearing points of lag screw) are visible at the distal medial edge.The counterparts were found on the lag screw in the areas with corresponding friction signs and worn edges.Bearing points are typical results of the interaction of the single products under load.Based on investigation the nail breakage is not linked to a deficiency of the device but is rather considered patient and / or procedure related.However, an exact root cause could not be determined.It could not be excluded that the implant had to absorb / transfer the whole loading because it was not relieved by increasing bone healing.Due to missing x-rays, it could not be determined whether reduction had been performed according to anatomical requirements and furthermore, if the implant positioning was according to anatomical requirements as well.Furthermore, as no further information such as x-rays, medical records or surgery reports was provided, a real medical statement was impossible.Finally, it could not be determined what range of weight bearing had been advised by the surgeon for the post-operative period.It could also not be determined if the patient had been compliant to the instructions.General aspects: the gamma3 nail is a temporary implant which inevitably will be subject of a fatigue fracture if the biomechanical stresses on the implant are too high or not considerably reduced during the period of implantation.During this period there is a race between bony consolidation / fracture healing and implant breakage as noted in the labelling of the product.Usually, a breakage is contributed by one or more deficits, e.G.Insufficient bone healing, product damage.Generally, the risk of a breakage will increase with the increase of load cycles and load level.Nail breakage in general has been experienced but does not present an unanticipated event in itself.Depending on load application, also, depending on the patient¿s post implant behaviour and depending on suitable anatomical reduction, depending on the kind of bone breakage, depending on the course of bone healing and other factors ¿ e.G.Increased post-operative activities ¿ a nail breakage can rather be classified as anticipated; specifically, if one or more contributing issues concurrence with each other.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.According to available information a deficiency of the nail was not verified.In case substantive information will become available in future that suggests otherwise we reserve the right to reopen the case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GAMMA3 - LONG NAIL KIT R1.5, TI, LEFT D11X420MM X 125DEG
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key22632285
Report Number0008031020-2025-01306
Device Sequence Number20831667
Product Code HSB
UDI-Device Identifier07613252274308
UDI-Public(01)07613252274308(17)271130(10)K0A9635
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K200869
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number35251420S
Device Lot NumberK0A9635
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/01/2025
Supplement Date Received by Manufacturer10/15/2025
Initial Report FDA Received Date07/28/2025
Supplement Report FDA Received Date11/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/02/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-