|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Updated fields: d8, d9, h2, h3, h6 and h11.The device was returned to the regional service center for inspection and the reported failure was confirmed.In addition, the following reportable malfunction was identified during the device evaluation: connector damage and adapter mount corrosion.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: regarding the customer¿s allegation, since the no video image output is cause was traced to component failure.And for the newly identified malfunction mentioned above, is cause not established.Should additional relevant information become available, a supplemental report will be submitted.
|