| Model Number |
0185 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that upon review of boston scientific latitude remote monitoring system stored data, we identified evidence that this lead exhibited high shock impedance measurements of greater than 125 ohms.No adverse patient effects have been reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that upon review of boston scientific latitude remote monitoring system stored data, we identified evidence that this lead exhibited high shock impedance measurements of greater than 125 ohms.No adverse patient effects have been reported.Additional information was received that this right ventricular (rv) lead exhibited a gradual rise in shock impedances since 2023.The field representative noted this patient did not pass in initial polarity.Ultimately, the device was explanted due to a normal generator change.During the change out procedure defibrillation threshold (dft) test was performed with initial polarity and triad configuration.The first attempt was unsuccessful in initial polarity and a second and third attempt were both successful, with in range shock impedance measurements.This rv lead remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific and physical analysis could not be conducted.Investigation of the available information/data determined that the observed gradual rise in lvsi measurements was likely associated with calcification, as there was no conclusive evidence of compromised physical or electrical integrity of the lead.This condition may develop over time and can be reflected in a gradual and sustained increase in shock impedance measurements.Boston scientific issued a field safety notice to physicians in (b)(6) 2025 regarding the potential for reliance defibrillation leads with expanded polytetrafluoroethylene (eptfe) coating to exhibit a pattern of gradually increasing lvsi measurements due to calcification, which may result in elevated high voltage shock impedance (hvsi) measurements and reduced shock efficacy.
|
| |
|
Event or Problem Description
|
|
It was reported that upon review of boston scientific latitude remote monitoring system stored data, we identified evidence that this lead exhibited high shock impedance measurements of greater than 125 ohms.No adverse patient effects have been reported.Additional information was received that this right ventricular (rv) lead exhibited a gradual rise in shock impedances since 2023.The field representative noted this patient did not pass in initial polarity.Ultimately, the device was explanted due to a normal generator change.During the change out procedure defibrillation threshold (dft) test was performed with initial polarity and triad configuration.The first attempt was unsuccessful in initial polarity and a second and third attempt were both successful, with in range shock impedance measurements.This rv lead remains in service.No adverse patient effects were reported.
|
| |
|