BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE G; IMPLANTABLE LEAD Back to Search Results
Model Number 0185
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that upon review of boston scientific latitude remote monitoring system stored data, we identified evidence that this lead exhibited high shock impedance measurements of greater than 125 ohms.No adverse patient effects have been reported.
 
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that upon review of boston scientific latitude remote monitoring system stored data, we identified evidence that this lead exhibited high shock impedance measurements of greater than 125 ohms.No adverse patient effects have been reported.Additional information was received that this right ventricular (rv) lead exhibited a gradual rise in shock impedances since 2023.The field representative noted this patient did not pass in initial polarity.Ultimately, the device was explanted due to a normal generator change.During the change out procedure defibrillation threshold (dft) test was performed with initial polarity and triad configuration.The first attempt was unsuccessful in initial polarity and a second and third attempt were both successful, with in range shock impedance measurements.This rv lead remains in service.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
This product has not been returned to boston scientific and physical analysis could not be conducted.Investigation of the available information/data determined that the observed gradual rise in lvsi measurements was likely associated with calcification, as there was no conclusive evidence of compromised physical or electrical integrity of the lead.This condition may develop over time and can be reflected in a gradual and sustained increase in shock impedance measurements.Boston scientific issued a field safety notice to physicians in (b)(6) 2025 regarding the potential for reliance defibrillation leads with expanded polytetrafluoroethylene (eptfe) coating to exhibit a pattern of gradually increasing lvsi measurements due to calcification, which may result in elevated high voltage shock impedance (hvsi) measurements and reduced shock efficacy.
 
Event or Problem Description
It was reported that upon review of boston scientific latitude remote monitoring system stored data, we identified evidence that this lead exhibited high shock impedance measurements of greater than 125 ohms.No adverse patient effects have been reported.Additional information was received that this right ventricular (rv) lead exhibited a gradual rise in shock impedances since 2023.The field representative noted this patient did not pass in initial polarity.Ultimately, the device was explanted due to a normal generator change.During the change out procedure defibrillation threshold (dft) test was performed with initial polarity and triad configuration.The first attempt was unsuccessful in initial polarity and a second and third attempt were both successful, with in range shock impedance measurements.This rv lead remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE G
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112 5798
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key22635569
Report Number2124215-2025-50675
Device Sequence Number18454407
Product Code LWS
UDI-Device Identifier00802526531194
UDI-Public(01)00802526531194(17)150115
Combination Product (Y/N)Y
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P910073
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall,Notification
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/15/2015
Device Model Number0185
Device Catalogue Number0185
Device Lot Number361913
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/10/2025
Supplement Date Received by Manufacturer07/30/2025
02/11/2026
Initial Report FDA Received Date07/28/2025
Supplement Report FDA Received Date08/11/2025
02/24/2026
Date Device Manufactured01/15/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age63 YR
Patient SexMale
-
-