It was reported that a patient underwent an unknown orthopedic surgery on an unknown date and barbed suture was used.During the procedure, it was reported by a sales rep that the suture was broken during operation.This event occurred when the doctor tried to suture the wound as usual and the suture was pulled, but the suture at the middle part suddenly was broken.It was not a control release needle.Additional suture was performed to complete the case.Further details are not provided from the hp.Affiliate reference number: (b)(4).No adverse patient consequences were reported.Additional information was requested.
|
|
Product complaint # (b)(4).H3 investigation summary: the product was returned to ethicon for evaluation.Two needle-suture pieces outside its packaging were received for analysis.Product code sxpp2b100.During the visual inspection of the suture, no breakage suture was observed.Even though the extreme was noted to be broken probably caused by a surgical instrument.Furthermore, body fluids were also observed on the strand.In addition, marks that appear to be caused by a surgical instrument were observed on the body needle.This product code sxpp2b100 contains an absorbable suture.Since the sample was received open, the functional test could not be performed due to undetermined exposure time to the environment.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: the instructions for use contained the following caution: as with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of surgical instruments, such as needle holders and forceps.The event described could not be confirmed.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.
|