| Catalog Number |
367341 |
| Medical Device Problem Code |
Difficult or Delayed Activation (2577)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when using one (1) bd vacutainer® push button blood collection set, the needle did not fully retract after pressing the button.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
D2b: additional medical device type: fpa.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary bd had not received any samples or photos for investigation.However, 30 retained samples underwent functional tests, all of which passed as they were activated properly with no retraction failure or defects observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: does not retract.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that when using one (1) bd vacutainer® push button blood collection set, the needle did not fully retract after pressing the button.No adverse impact was reported regarding the patient or user.
|
| |
|