• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO., (BD) BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET; BLOOD SPECIMEN COLLECTION DEVICE; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO., (BD) BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET; BLOOD SPECIMEN COLLECTION DEVICE; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 367341
Medical Device Problem Code Difficult or Delayed Activation (2577)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/08/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when using one (1) bd vacutainer® push button blood collection set, the needle did not fully retract after pressing the button.No adverse impact was reported regarding the patient or user.
 
Additional Manufacturer Narrative
D2b: additional medical device type: fpa.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
Investigation summary bd had not received any samples or photos for investigation.However, 30 retained samples underwent functional tests, all of which passed as they were activated properly with no retraction failure or defects observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: does not retract.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
 
Event or Problem Description
It was reported that when using one (1) bd vacutainer® push button blood collection set, the needle did not fully retract after pressing the button.No adverse impact was reported regarding the patient or user.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE; INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
matt j. connor
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key22636707
Report Number1024879-2025-01036
Device Sequence Number9074319
Product Code JKA
UDI-Device Identifier00382903673414
UDI-Public(01)00382903673414(17)260430(10)4149141
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K220212
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date04/30/2026
Device Catalogue Number367341
Device Lot Number4149141
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2025
Supplement Date Received by Manufacturer09/18/2025
Initial Report FDA Received Date07/28/2025
Supplement Report FDA Received Date10/16/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/28/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-