DEPUY ORTHOPAEDICS INC US ARTICUL/EZE BALL 32 +5 BR; ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS Back to Search Results
Catalog Number 136522000
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Pain (1994); Joint Dislocation (2374)
Date of Event 01/10/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #(b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient complained of pain in operative side hip, and surgeon found that patient was chronically dislocating.During revision surgery it was determined that liner had excessive wear which allowed the dislocations to occur.Head was removed from stem and cup liner was exchanged.Old liner and new liner are both competitor products.Nothing found wrong or defective with our head.Doi: (b)(6) 2020, dor: (b)(6) 2025, affected side: right hip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICUL/EZE BALL 32 +5 BR
Common Device Name
ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JJMSZ PLANT FOR STK. & N-STK.
no. 299, changyang street
suzhou industrial park
suzhou, jiangsu 21512 6
CH   215126
Manufacturer Contact
elba bello
700 orthpaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key22636750
Report Number1818910-2025-13068
Device Sequence Number9251513
Product Code JDI
UDI-Device Identifier10603295033189
UDI-Public(01)10603295033189(17)240630(10)D19070528
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K883460
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2024
Device Catalogue Number136522000
Device Lot NumberD19070528
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/10/2025
Supplement Date Received by Manufacturer07/31/2025
Initial Report FDA Received Date07/28/2025
Supplement Report FDA Received Date08/01/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/08/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age78 YR
Patient SexMale
-
-