Model Number 7300TFX
Medical Device Problem Codes Calcified (1077); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Dyspnea (1816); Fatigue (1849); Insufficient Information (4580)
Date of Event 06/27/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H10: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
It was learned through implant patient registry that a 31mm 7300tfx mitral valve was explanted after an implant duration of 8 years and 5 months due to unknown reason.The valve was replaced with a 33mm 11400m mitral valve.
 
Event or Problem Description
It was learned through implant patient registry and investigation that a 31mm 7300tfx mitral valve was explanted after an implant duration of 8 years and 5 months due to calcification and mod-severe stenosis with high mg.The patient presented with shortness of breath, fatigue.The valve was replaced with a 33mm 11400m mitral valve.The patient was discharged on pod #6.
 
Additional Manufacturer Narrative
H10: additional manufacturer narrative: updated sections b5, b7, h6 (device code, clinical code, type of investigation).
 
Additional Manufacturer Narrative
Svd is a logical and expected consequence of the chemical, mechanical, and immunological processes associated with bhv implantation.The common endpoint of all these factors is calcification and degradation.Svd is an acquired condition, intrinsic to bioprosthetic valves, characterized by deterioration of the leaflets or supporting structures, leading to thickening, calcification, tearing, or disruption of the prosthetic valve materials.These changes result in valvular dysfunction manifesting as stenosis and/or regurgitation with a consequent drop in hemodynamic efficacy of the valve.Despite the advancements in the heart valve industry, svd commonly begins 7- 8 years after implantation, with a considerably increased rate in patients with pertinent comorbidities and/or an implant duration of greater than 10 years.Prior investigations and extensive literature review have shown that svd is predominantly related to patient factors and implant duration rather than to manufacturing issues.Events related to device design, manufacturing, or use error tend to manifest at an earlier implant duration.Calcification plays a major role in the failure of bioprosthetic heart valves.Calcification of valves occurs as a progressive, time-dependent process.Tissue valve calcification is initiated primarily within residual cells that have been devitalized.Initial calcification deposits eventually enlarge and grow into a mass, which stiffen and weaken the tissue and thereby cause the prosthesis to malfunction.The mineralization of a biomaterial is generally enhanced at the sites of intense mechanical deformations generated by motion, such as the points of flexion in heart valves.Ultimately, the result of calcification is valve failure due to tearing or stenosis.Many factors contribute to the onset and propagation of calcification.These include patient factors (age, disease state, pharmacological intervention, etc.), mechanical stress related to the valve's hemodynamic performance, and glutaraldehyde fixation of tissue.Of these, the fixation process is a relatively minor contributor to calcification for edwards tissue valves due to anti-calcification treatments during manufacturing.Though numerous studies have been conducted on preventive calcification strategies in bioprosthetic heart valves, the causes of calcification are not fully understood and there are still no mechanisms or medical therapies which fully prevent bioprosthesis from calcifying.The root cause of this event cannot be conclusively determined with the available information.The explanted device is not available for evaluation.However, the calcification observed in this case was most likely due to a progression of the patients underlying valvular disease pathology combined with the patients other underlying risk factors, including an implant duration of 7 or more years and hyperlipidemia the device history record (dhr) could not be reviewed, as the device serial number was not provided.However, the reported event does not allege a malfunction that could be related to a manufacturing deficiency and/or one was not confirmed through investigation.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL EASE PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
irvine, CA 92614
9492500222
MDR Report Key22636985
Report Number2015691-2025-06113
Device Sequence Number9251519
Product Code DYE
UDI-Device Identifier00690103186284
UDI-Public(01)00690103186284(17)200217
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/17/2020
Device Model Number7300TFX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2025
Supplement Date Received by Manufacturer07/31/2025
08/22/2025
Initial Report FDA Received Date07/28/2025
Supplement Report FDA Received Date08/21/2025
09/04/2025
Date Device Manufactured02/18/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization; Life Threatening;
Patient Age59 YR
Patient SexMale
-
-