|
(b)(4) date sent: 07/28/2025 d4: batch #538d41.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one pse45a device was returned with no apparent damage and with an ecr45b reload present.The reload was received unfired, with 1 double driver missing, making the reload non-functional.In addition, the reload pan was noted to be partially dislodged from the left proximal side.The device was tested for functionality in the straight position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples met the staple release criteria.As part of our quality process, the manufacturing records of this batch number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached on what may have caused the reload pan to dislodge.A manufacturing record evaluation was performed for the finished device pse45a lot number a97v68 and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number 538d41, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|