Model Number V+C, ENGLISH
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/03/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H6: ventec will perform an evaluation of the device.A follow-up report will be submitted when the investigation is complete as defined by 21 cfr 803.56.
 
Event or Problem Description
It was reported to ventec that the device was delivering higher than set peep (positive end expiratory pressure).There was patient involvement associated with the reported event; however, there were no reports of patient harm as a result of the reported issue.Ventec has made multiple attempts to obtain additional information about the patient, but no response has been received.
 
Additional Manufacturer Narrative
H6: ventec has made multiple attempts to contact the initial reporter asking if the device will be returned for an evaluation, however, no response has been received.In the event that additional information is received, ventec will submit a follow-up report as defined by 21 cfr 803.56.Ventec performed a review of the device's manufacturing records which showed all requirements were met.Final test specifications were acceptable.No non-conformities or anomalies were found related to this complaint when reviewing the device history record.Trend analysis and risk analysis were considered acceptable.The device was not returned to ventec for an evaluation.The cause of the reported issue could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOCSN
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VENTEC LIFE SYSTEMS, INC
22002 26th ave se
bothell WA 98021
Manufacturer (Section G)
VENTEC LIFE SYSTEMS, INC
22002 26th ave se
bothell WA 98021
Manufacturer Contact
elizabeth gilbert
22002 26th ave se
bothell, WA 98021
4256861765
MDR Report Key22638287
Report Number3013095415-2025-00687
Device Sequence Number9563923
Product Code CBK
UDI-Device Identifier00855573007730
UDI-Public(01)00855573007730(11)240920
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K162877
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberV+C, ENGLISH
Device Catalogue NumberPRT-01098-000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/03/2025
Supplement Date Received by Manufacturer04/01/2026
Initial Report FDA Received Date07/28/2025
Supplement Report FDA Received Date04/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/20/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-