H3: product analysis of product: 5330008, lotno:id20l016 visual and optical inspection confirmed the laser weld on one side of the handle has broken.There is some material deformation around the laser weld.The damage appears to be from overload as the mechanism of failure.E1: initial reporter is unknown.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|