| Catalog Number |
381447 |
| Medical Device Problem Code |
Retraction Problem (1536)
|
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
| Date of Event |
07/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3.The date received by manufacturer has been used for this field.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd insyte autoguard needle partially retracted.The following information was provided by the initial reporter: a colleague at ***** used this device and was injured when the iv needle did not fully retract.
|
| |
|
Event or Problem Description
|
|
No new information.
|
| |
|
Additional Manufacturer Narrative
|
|
Since no samples displaying the condition reported were available for examination, we were unable to fully investigate this incident.However, the reported defect of needle retraction failure is confirmed since a trend has been identified for the reported failure mode and corrective actions have been initiated to investigate this type of incident and identify the root cause.We would be very interested in examining product that does not meet your expectations and our quality standards.Should you again experience any problems with our product we would appreciate the opportunity to conduct a thorough analysis.Examination of the actual product involved may provide clarification as to the cause for the reported failure.A device history record review showed no non-conformances associated with this issue during the production of this batch.
|
| |
|
Search Alerts/Recalls
|