Model Number GM-324
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D10 concomitant products: gm-324, gm-324 biosyn 3-0 vio 75cm cv25 x36 (lot#:d5b3264y), gm-324, gm-324 biosyn 3-0 vio 75cm cv25 x36 (lot#:d5b3264y), gm-324, gm-324 biosyn 3-0 vio 75cm cv25 x36 (lot#:d5b3264y).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during a veterinary procedure, when withdrawing from the retainer and preparing for suturing, the thread broke.There were four sutures used and had the same issue.A new suture was used to resolve the issue.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted three sutures on the returned product, though the needles were not included.The sutures were broken, with lengths measuring 21 1/2 inches, 21 inches, and 25 1/8 inches, using a calibrated aluminum rule.It was reported that the thread was broken.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: progressive loss of tensile strength and eventual absorption of the absorbable sutures occurs by means of hydrolysis.This loss is triggered when exposed to the environment or clinical application.Damage during use may result in inadequate sample length to perform testing and/or deformation of the suture contributing to a loss in tensile strength.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOSYN
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key22639774
Report Number9612501-2025-02051
Device Sequence Number9538394
Product Code GAM
UDI-Device Identifier20884521032719
UDI-Public20884521032719
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K000037
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGM-324
Device Catalogue NumberGM-324
Device Lot NumberD5B3264Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2025
Supplement Date Received by Manufacturer09/10/2025
Initial Report FDA Received Date07/28/2025
Supplement Report FDA Received Date09/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NOTE ON H11
Patient SexUnknown
-
-