| Catalog Number |
1810MS |
| Medical Device Problem Codes |
Difficult to Remove (1528); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records will be performed.The device has not been returned to the manufacturer for evaluation.However, a photo and video were provided for review.The investigation of the reported event is currently underway.Section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025 a patient underwent an ultrasound guided breast biopsy procedure through hard density tissue using mission kit.During the procedure the needle was bent and difficult to remove from the patient.The procedure was completed by another device.There was no patient reported injury.
|
| |
|
Event or Problem Description
|
|
(b)(6) 2025 a patient underwent an ultrasound guided breast biopsy through hard density tissue using mission kit.During the procedure, the needle was bent and difficult to remove from the patient.The procedure was completed by another device.There was no patient reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.One electronic photo was provided and reviewed.The photo shows one mission device with 20 mm configuration placed on a flat surface.The needle of the device was noted to be bent.No other visual anomalies were noted.One electronic video was provided and reviewed.The video shows one mission device was held by the healthcare professional in 20 mm configuration.He rotates the device, and the needle of the device was noted to be bent.No other visual anomalies were noted.Based on the photo and video review the reported bent can be confirmed for one device.Therefore, the investigation is confirmed for the reported bent as a bent was noted on the needle of one device.However, the investigation is also inconclusive for the reported difficult to remove for first device as absence of supporting evidence.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.D4 (unique identifier (udi) #), e1, g3, h6 (method, result, conclusion).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|