| Model Number |
8422500 |
| Medical Device Problem Codes |
No Display/Image (1183); Gas Output Problem (1266); Decrease in Pressure (1490); Intermittent Communication Failure (4038)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported where a v800 ventilator display turned off on a patient during ventilation.The patient on the ventilator continued to receive breaths and patient was not harmed.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation has just started; results will be provided in a follow-up report.
|
| |
|
Additional Manufacturer Narrative
|
|
The log file of the affected device was provided for the investigation.The analysis of the log file could confirm that the device performed a restart on the reported date of event.The device restart was caused by a potential temporary malfunction of pba m48.3.The pba m48.3 and the micro sd card were requested for further investigation but have not been provide.As a result, the underlying root cause could not be clarified.As a safety feature of the system, the safety software analyses and verifies proper function of the device.In case of a detected deviation regarding operation of the ventilation unit, the safety software triggers a synchronized restart of the ventilation unit and the display unit (ecd) in order to reset the system to a specified state.During restart sequence the ventilation is temporarily interrupted, and the inspiratory valve is opened to ambient allowing the patient for spontaneous breathing.The deviation will be indicated by activated auxiliary auditory alarm.Single restart sequence of the ventilation unit takes up to 8 seconds until evita v800 automatically resumes the ventilation with the latest settings.The restart sequence of the ecd may take up to a maximum of 1 minute.In the meantime, the user can observe the already resumed ventilation and safety-relevant parameters such as fio2 concentration, minute volume and airway pressure in the oled-display of the ventilation unit.Finally, the alarm message "ventilation unit restarted" will be prompted on the screen with accompanying auditory alarm and the auxiliary auditory alarm ceases automatically.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|
| |
|
Event or Problem Description
|
|
It was reported where a v800 ventilator display turned off on a patient during ventilation.The patient on the ventilator continued to receive breaths and patient was not harmed.
|
| |
|