DRÄGERWERK AG & CO. KGAA EVITA V800; VENTILATORS, INTENSIVE CARE Back to Search Results
Model Number 8422500
Medical Device Problem Codes No Display/Image (1183); Gas Output Problem (1266); Decrease in Pressure (1490); Intermittent Communication Failure (4038)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/15/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported where a v800 ventilator display turned off on a patient during ventilation.The patient on the ventilator continued to receive breaths and patient was not harmed.
 
Additional Manufacturer Narrative
The investigation has just started; results will be provided in a follow-up report.
 
Additional Manufacturer Narrative
The log file of the affected device was provided for the investigation.The analysis of the log file could confirm that the device performed a restart on the reported date of event.The device restart was caused by a potential temporary malfunction of pba m48.3.The pba m48.3 and the micro sd card were requested for further investigation but have not been provide.As a result, the underlying root cause could not be clarified.As a safety feature of the system, the safety software analyses and verifies proper function of the device.In case of a detected deviation regarding operation of the ventilation unit, the safety software triggers a synchronized restart of the ventilation unit and the display unit (ecd) in order to reset the system to a specified state.During restart sequence the ventilation is temporarily interrupted, and the inspiratory valve is opened to ambient allowing the patient for spontaneous breathing.The deviation will be indicated by activated auxiliary auditory alarm.Single restart sequence of the ventilation unit takes up to 8 seconds until evita v800 automatically resumes the ventilation with the latest settings.The restart sequence of the ecd may take up to a maximum of 1 minute.In the meantime, the user can observe the already resumed ventilation and safety-relevant parameters such as fio2 concentration, minute volume and airway pressure in the oled-display of the ventilation unit.Finally, the alarm message "ventilation unit restarted" will be prompted on the screen with accompanying auditory alarm and the auxiliary auditory alarm ceases automatically.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
 
Event or Problem Description
It was reported where a v800 ventilator display turned off on a patient during ventilation.The patient on the ventilator continued to receive breaths and patient was not harmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVITA V800
Common Device Name
VENTILATORS, INTENSIVE CARE
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23558
GM   23558
Manufacturer Contact
moislinger allee 53-55
lübeck 23558
4518822868
MDR Report Key22641056
Report Number9611500-2025-00416
Device Sequence Number5418995
Product Code CBK
UDI-Device Identifier04048675542148
UDI-Public(01)04048675542148(11)231006(93)8422500-08
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K222024
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 03/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8422500
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/01/2025
Supplement Date Received by Manufacturer03/05/2026
Initial Report FDA Received Date07/29/2025
Supplement Report FDA Received Date03/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/06/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NA.; NA.