Model Number 6688600
Medical Device Problem Code Failure of Device to Self-Test (2937)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/02/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
On-site investigation was performed by our field service engineer.The claimed issue was reproduced during troubleshooting.According to received information in service report, the replacement of the expiratory valve coil solved the issue.The ventilator passed all functional and safety tests and was cleared for clinical use.The movable axis of the expiratory valve coil controls the opening of the expiratory valve by pushing the valve membrane into desired position.The power supply to the coil is regulated so that the remaining pressure in the patient system, towards the end of the expiration time, is kept on the peep level according to user interface setting.Defective expiratory valve coil may lead to stop of ventilation.The reported issue was confirmed in provided device's logs.The claimed part was not provided for further analysis.The cause of the claimed issue in this customer product complaint has been determined.The issue was related to expiratory valve coil.
 
Event or Problem Description
It was reported that the ventilator failed pressure transducer test during pre-use check.There was no patient involvement.Manufacturer's ref.#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BASE UNIT SERVO-N
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
caroline kabbabe
roentgenvagen 2
solna 
MDR Report Key22641225
Report Number8010042-2025-0001182
Device Sequence Number5194054
Product Code CBK
UDI-Device Identifier07325710001042
UDI-Public(01)07325710001042(11)161209
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K201874
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6688600
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/17/2025
Initial Report FDA Received Date07/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/09/2016
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1