| Model Number |
6688600 |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On-site investigation was performed by our field service engineer.The claimed issue was reproduced during troubleshooting.According to received information in service report, the replacement of the expiratory valve coil solved the issue.The ventilator passed all functional and safety tests and was cleared for clinical use.The movable axis of the expiratory valve coil controls the opening of the expiratory valve by pushing the valve membrane into desired position.The power supply to the coil is regulated so that the remaining pressure in the patient system, towards the end of the expiration time, is kept on the peep level according to user interface setting.Defective expiratory valve coil may lead to stop of ventilation.The reported issue was confirmed in provided device's logs.The claimed part was not provided for further analysis.The cause of the claimed issue in this customer product complaint has been determined.The issue was related to expiratory valve coil.
|
| |
|
Event or Problem Description
|
|
It was reported that the ventilator failed pressure transducer test during pre-use check.There was no patient involvement.Manufacturer's ref.#: (b)(4).
|
| |
|