ZIMMER BIOMET, INC. VANGUARD POSTERIOR STABILIZED TIBIAL BEARING 10MM X 71/75MM; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Unspecified Infection (1930); Insufficient Information (4580)
Date of Event 07/07/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4) d10 - concomitant devices - unknown vanguard femoral component catalog #: ni lot #: ni, unknown biomet tibial component catalog #: ni lot #: ni.G2 - report source - foreign: event occurred in australia.The complainant has not yet indicated whether the product will be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient underwent a knee arthroplasty revision of the tibial bearing to address unknown complications post-operatively.Attempts have been made, however, no additional information is available.
 
Event or Problem Description
It was reported that the patient underwent a left knee revision of the tibial bearing due to infection approximately eight (8) years following knee arthroplasty.Attempts have been made, however, no additional information is available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: a1, a2, a3, b4, b5, d1, d4, d6a, d10, g3, g4, g6, h2, h4, h6, h11.D10 - concomitant devices - biomet polished finned tibial tray 71mm catalog #: 141253 lot #: 2016090409, biomet series a 3 peg thin patella 37mm x 8.6mm catalog #: 184788 lot #: 160430, vanguard posterior stabilized interlok femoral component left 62.5mm catalog #: 183126 lot #: 924160.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, g3, g6, h2, h6, h11.The reported event could not be confirmed as medical records were not provided.The device history records were not reviewed as the event was determined to be unrelated to the implanted device.Sterile certifications were reviewed and found to be conforming with no applicable deviations.The device was verified to have gone through acceptable sterilization processes following iso/aami/astm & eu published guidelines.There are multiple factors that may contribute to an infection that are outside the control of zimmer biomet, such as external factors that may include hospital/surgical environment, provider-related risk factors, and/or patient comorbidities/risk factors.As there are no indications of a product or process issue affecting implant safety or effectiveness, the implanted products are not identified as the source or contributing to the reported infection.The root cause was determined to be traced to non-device related factors.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VANGUARD POSTERIOR STABILIZED TIBIAL BEARING 10MM X 71/75MM
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key22642267
Report Number0001825034-2025-02297
Device Sequence Number18313270
Product Code JWH
UDI-Device Identifier00880304271937
UDI-Public(01)00880304271937(17)220506(10)749170
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K171054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/06/2022
Device Model NumberN/A
Device Catalogue Number183640
Device Lot Number749170
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2025
Supplement Date Received by Manufacturer08/15/2025
10/09/2025
Initial Report FDA Received Date07/29/2025
Supplement Report FDA Received Date09/05/2025
10/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/06/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
-
-