• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA MECTACER BIOLOX DELTA FEMORAL BALL HEAD; MECTACER MECTACER HEAD BIOLOX DELTA DIA.36 12/14-S

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA MECTACER BIOLOX DELTA FEMORAL BALL HEAD; MECTACER MECTACER HEAD BIOLOX DELTA DIA.36 12/14-S Back to Search Results
Model Number 01.29.208
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Medical device site infection (4845)
Date of Event 07/12/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 23 july 2025.Ball heads: mectacer 01.29.208 mectacer head biolox delta dia.36 12/14-s (k112115) lot.2432821: (b)(4) items manufactured and released on 10-dec-2024.Expiration date: 26-nov-2029.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Liner: mpact 01.32.3644hct flat pe hc liner ø36/e (k103721) lot.2436032: (b)(4) items manufactured and released on 04-march-2025.Expiration date: 11-feb-2030.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Conclusions: infection is a known possible complication of any surgery.Although no root cause can be established, there is no indication that any issue with the device may have caused or contributed to the event, and the document review does not indicate any potential manufacturing related cause.
 
Event or Problem Description
At about 1 month from the primary surgery, the patient came in due to signs of an infection and the pathogen is unknown.The surgeon performed a washout and revised the head and liner.The surgery was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MECTACER BIOLOX DELTA FEMORAL BALL HEAD
Common Device Name
MECTACER MECTACER HEAD BIOLOX DELTA DIA.36 12/14-S
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key22643027
Report Number3005180920-2025-00732
Device Sequence Number8632304
Product Code LZO
UDI-Device Identifier07630030809194
UDI-Public07630030809194
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K112115
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number01.29.208
Device Lot Number2432821
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/12/2025
Initial Report FDA Received Date07/29/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/10/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexFemale
Patient Weight113 KG
Patient RaceWhite
-
-