| Model Number |
01.18.435 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Medical device site infection (4845)
|
| Date of Event |
07/12/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on 23 july 2025.Stem: amistem-p collared 01.18.435 amistem-p collared std stem size 5 (k173794) lot.2413784: (b)(4) items manufactured and released on 25-sept -2024.Expiration date: 11-sept-2029.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Ball heads: mectacer 01.29.210 mectacer head biolox delta dia.36 12/14-l (k112115) lot.2505494: (b)(4) items manufactured and released on 17-march-2025.Expiration date: 02-march-2030.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Liner: mpact 01.32.3648hct flat pe hc liner ø36/f (k103721) lot.2435355: (b)(4) items manufactured and released on 04-march-2025.Expiration date: 13-feb-2030.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Conclusions: infection is a known possible complication of any surgery.Although no root cause can be established, there is no indication that any issue with the device may have caused or contributed to the event, and the document review does not indicate any potential manufacturing related cause.
|
| |
|
Event or Problem Description
|
|
At about 1 month from the primary surgery, the patient came in due to signs of an infection, and the pathogen is unknown.The surgeon performed a washout and revised the stem, head, and liner.The surgery was completed successfully.
|
| |
|
Search Alerts/Recalls
|