• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. CLAREON IOL WITH THE AUTONOME DELIVERY SYSTEM; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. CLAREON IOL WITH THE AUTONOME DELIVERY SYSTEM; INTRAOCULAR LENS Back to Search Results
Model Number CNA0T0
Medical Device Problem Codes Device Difficult to Setup or Prepare (1487); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2025
Type of Reportable Event Malfunction
Event or Problem Description
A non-healthcare professional reported that during intraocular lens (iol) implantation surgery, the haptic was not folded properly and lens advanced with difficulty.The lens was examined under a microscope and was seen to be damaged.Additional information has been received stating that there was no patient contact, the device tip didn't come in contact with the patient and the procedure was completed on the same day.
 
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information was provided in d.9., h.3., h.6., h.11.The product was returned for analysis.The device was returned loose in the product carton.The lock-out assembly has been removed.Viscoelastic is observed in the device.The plunger has been advanced at the wound guard and is advanced slightly under the iol (intraocular lens).The iol is advanced to the wound guard and is damaged.We are unable to determine the root cause for the reported complaint.The reporter stated: "the haptic was not folded properly", misfolding of the haptic event can occur: if the lens has become misaligned in the lens bay during the manufacturing process.Due to the use of a non-qualified viscoelastic.Material properties of non-qualified ovds (ophthalmic viscosurgical device) may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.If inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly.If the operating room temperature is too high (more than 23 degree celsius / 73-degree fahrenheit) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.Any of the above listed causes alone, or in combination, may create the reported event.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON IOL WITH THE AUTONOME DELIVERY SYSTEM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22643760
Report Number9612169-2025-01498
Device Sequence Number8678059
Product Code HQL
UDI-Device Identifier00380652393805
UDI-Public(01)00380652393805(17)271127(11)241128(10)25955144
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCNA0T0
Device Lot Number25955144
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/07/2025
Supplement Date Received by Manufacturer09/29/2025
Initial Report FDA Received Date07/29/2025
Supplement Report FDA Received Date10/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/28/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DUOVISC VISCOELASTIC SYSTEM
Patient Age58 YR
Patient SexFemale
-
-