Event or Problem Description
A trilogy 100 ventilator was returned to a third-party service center for service.There was no reported patient involvement.During the evaluation at the third-party service center, the device was visually inspected, and foam particles were observed.The device was remediated, and the foam insulation was replaced to resolve the issue.Additional findings not related to the malfunction/issue missing rubber feet, handle o-rings, and the right button of the keypad's coating was worn.There was dust found on the blower box by the third-party service technician.The customer requested that the unit be returned unrepaired.No further investigation is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 200
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key22648269
Report Number2518422-2025-108356
Device Sequence Number12965734
Product Code CBK
UDI-Device Identifier00606959026360
UDI-Public(01)00606959026360
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryIT
PMA/510(K) Number
K083526
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 07/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number1032800B
Device Catalogue Number1032800B
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/17/2025
Initial Report FDA Received Date07/29/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/12/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1956-2021
Patient Sequence Number1
Patient SexUnknown
-
-