| Catalog Number |
362760 |
| Medical Device Problem Code |
Coagulation in Device or Device Ingredient (1096)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported during use of bd vacutainer® cpt¿ cell preparation tube with sodium citrate, an unspecified number of samples exhibit clumped/sticky mononuclear cells after top layer is removed from tubes and placed in a conical tube.There was no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd had not received any samples or photos for investigation.A total of 50 retained samples were inspected, with no issues being identified.Complaints for sample quality are under statistical control for the month of july 2025.At this time, further testing is not indicated.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: sample quality-poor plasma.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.Bd quality will continue to monitor sample quality complaints.
|
| |
|
Event or Problem Description
|
|
It was reported during use of bd vacutainer® cpt¿ cell preparation tube with sodium citrate, an unspecified number of samples exhibit clumped/sticky mononuclear cells after top layer is removed from tubes and placed in a conical tube.There was no health impact or consequences reported.
|
| |
|
Search Alerts/Recalls
|