• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE SG; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE SG; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number 0182
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/27/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that upon review of boston scientific latitude remote monitoring system stored data, we identified evidence that this lead exhibited high shock impedance measurements of greater than 125 ohms.No adverse patient effects have been reported.
 
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that upon review of boston scientific latitude remote monitoring system stored data, we identified evidence that this lead exhibited high shock impedance measurements of greater than 125 ohms.No adverse patient effects have been reported.Additional information was received that this right ventricular (rv) lead was surgically abandoned.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE SG
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key22649183
Report Number2124215-2025-51490
Device Sequence Number8798784
Product Code LWS
UDI-Device Identifier00802526531163
UDI-Public(01)00802526531163(17)150213
Combination Product (Y/N)Y
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P910073
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/13/2015
Device Model Number0182
Device Catalogue Number0182
Device Lot Number300912
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/10/2025
Supplement Date Received by Manufacturer10/15/2025
Initial Report FDA Received Date07/29/2025
Supplement Report FDA Received Date10/20/2025
Date Device Manufactured02/13/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age68 YR
Patient SexMale
-
-