ALLERGAN (COSTA RICA) INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number SCM-405
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported right side capsular contracture baker grade iii with "marked distortion and wrinkling".Device was explanted and replaced.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist abbvie in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii.
 
Additional Manufacturer Narrative
Device evaluation: based on the product analysis performed, the assessments of the complaint codes are: ¿ device wrinkling: creases observed on the device.¿ capsular contracture: unable to observe as it is not related to the manufacturing process.As per the investigation procedure, deformation and wear abrasion were observed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.Additional, changed, and/or corrected data: a4, a5, b5, d9, e1, h3, h6.
 
Event or Problem Description
Healthcare professional reported right side "no complaint against the device".Healthcare professional later reported capsular contracture baker grade iii with "marked distortion and wrinkling".Healthcare professional additionally reported:"any creases".Device has been explanted and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
8479366324
MDR Report Key22652671
Report Number9617229-2025-12745
Device Sequence Number8678411
Product Code FTR
UDI-Device Identifier10888628029927
UDI-Public(01)10888628029927(10)1209243(11)231003(17)280903
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberSCM-405
Device Lot Number1209243
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/02/2025
Supplement Date Received by Manufacturer08/13/2025
Initial Report FDA Received Date07/29/2025
Supplement Report FDA Received Date09/08/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient RaceWhite
-
-