| Model Number |
8884720221 |
| Medical Device Problem Code |
Unintended Deflation (4061)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Event or Problem Description
|
|
The customer reported that upon opening the gtt and performing the balloon test, it was full of punctures.The customer also states that the product packaging was intact, with no signs of damage or poor storage that could have caused the balloon to become punctured.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.Prior to a lot¿s release, the lot must be deemed acceptable by-passing inspections that are based on a valid sampling plan.During production, inspectors routinely examine a statistical sample both physically and visually.There were no samples returned for evaluation; therefore, the affected device could not be evaluated to confirm the reported failure mode.As such, a root cause could not be determined at this time.However, a supplier corrective action request has been sent to the supplier to further investigate and address the reported condition.We will continue to monitor related reports to determine if additional actions are necessary.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was received for investigation.The device was inspected, and the reported condition was observed.A supplier corrective action request has been sent to the supplier to further investigate and address the reported condition.We will continue to monitor related reports to determine if additional actions are necessary.
|
| |
|