• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number ACU0T0
Medical Device Problem Code Device Difficult to Setup or Prepare (1487)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed, and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non-healthcare professional reported that during an intraocular lens implant procedure the introducer did not slide on top of the lens.The surgery was completed and there was no patient harm.Additional information has been requested.
 
Additional Manufacturer Narrative
Additional information was provided in d.9., h.3., h.6.And h.11.The device with the lens was returned.The plunger lock and lens stop have been removed.The plunger was oriented correctly.No viscoelastic was observed in the device.The plunger was advanced under the lens to mid nozzle.The lens was located on top of the plunger still in the loading area.A device history record review and a nonconformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.A qualified viscoelastic was indicated.It is unknown what is meant by introducer did not slide on top of the lens.A plunger underride was observed.The root cause appears be related to a failure to follow the instructions for use (ifu).No viscoelastic was observed in the device.The ifu instructs: fill the device until viscoelastic can be observed flowing to the line on the nozzle tip.This will require approximately 0.2 ml of viscoelastic.If inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly or become stuck in the device.Plunger underride may occur due to rapid advancement faster than the ifu recommend rate.If inadequate viscoelastic is placed in the device, this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly or become stuck in the device allowing the plunger to underride the lens.If the operating room temperature is too high ( more than 23 degree c or 73 degree f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.Any of the above listed causes alone, or in combination, may create the reported event.Due diligence has been performed in an attempt to obtain further information on this event.Information was provided the reporter declined to provide additional information.No further information was provided.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM
Common Device Name
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22662885
Report Number1119421-2025-02090
Device Sequence Number12645782
Product Code KYB
UDI-Device Identifier00380652395335
UDI-Public(01)00380652395335(17)270813(11)240814(10)15906466
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberACU0T0
Device Lot Number15906466
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/07/2025
Supplement Date Received by Manufacturer08/27/2025
Initial Report FDA Received Date07/30/2025
Supplement Report FDA Received Date09/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/14/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PROVISC OVD
Patient SexUnknown
-
-