| Model Number |
24970A |
| Medical Device Problem Codes |
Computer Software Problem (1112); Communication or Transmission Problem (2896); Computer Operating System Problem (2898)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/24/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during an implant procedure, after connection of leads, the mobile programmer froze when attempting the impedance check.It was noted that the screen was unresponsive and when attempting to navigate away and back the screen went white.It was further noted that complete telemetry was not maintained with the implantable device throughout the session.Troubleshooting steps were taken to include force closing and restarting the application.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: performance data collected from the mobile programmer was received and analyzed.Analysis of the service logs found the customer comment that the mobile programmer was unresponsive/froze and telemetry was not maintained was confirmed.Analysis of the service logs found the customer comment that the mobile programmer screen went white was confirmed.Correction: d.2.Product code common device name medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|